• DocumentCode
    3563777
  • Title

    A testing procedure at wafer level for micro electrodes using a gel

  • Author

    Kindlundh, M.G. ; Norlin, P.

  • Author_Institution
    Acreo AB, Kista, Sweden
  • Volume
    1
  • fYear
    2003
  • Firstpage
    428
  • Abstract
    We demonstrate a novel 3-point impedance measurement procedure for microelectrodes. The concept is based on the use of a saline gel and enables relatively fast measurement at wafer level. In this study, measurements were first carried out with a saline gel as electrolyte, and then compared to a reference measurement with a conventional saline solution as electrolyte. The impedance as function of frequency was investigated for two different electrode sizes, and showed good agreement between both electrolytes. The testing procedure described in this paper can be applied to e.g. biomedical microsystems that must interact with a wet environment.
  • Keywords
    biomedical electrodes; electric impedance measurement; gels; microelectrodes; 3-point impedance measurement; biomedical microsystems; electrolyte; microelectrodes; saline gel; saline solution; wafer level; Biomedical measurements; Biosensors; Costs; Electrodes; Impedance; Microelectrodes; Packaging; Probes; Quality control; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    TRANSDUCERS, Solid-State Sensors, Actuators and Microsystems, 12th International Conference on, 2003
  • Print_ISBN
    0-7803-7731-1
  • Type

    conf

  • DOI
    10.1109/SENSOR.2003.1215345
  • Filename
    1215345