DocumentCode :
3565396
Title :
Comparative study of electrocardiogram QRS complex detection algorithm on Field Programmable Gate Array platform
Author :
Hashim, Amin ; Chia Yee Ooi ; Bakhteri, Rabia ; Yuan Wen Hau
Author_Institution :
Fac. of Electr. Eng., Univ. Teknol. Malaysia, Skudai, Malaysia
fYear :
2014
Firstpage :
241
Lastpage :
246
Abstract :
Nowadays, many people suffer from heart problems and hence the demand of inexpensive and efficient electrocardiogram (ECG) for frequent heart monitoring is becoming crucial. To make the ECG device portable, cost-effective and light-weight, an alternative of deploying an ECG system is on Field Programmable Gate Array (FPGA). It is also important to choose suitable algorithm that optimize in terms of feature extraction accuracy and computation time. This paper implements and compares two methods of ECG QRS detection in terms of Pan and Tompkins algorithm and Derivative-based Method on FPGA platform as embedded software computation. The Derivative-based Method is modified from fixed threshold to adaptive threshold to increase robustness in real time QRS detection. The input are 48 records of 30 minutes, total 24 hours ECG data obtained from MIT-BIH database as standard database for performance benchmarking. Both algorithm yield a different outcome in terms of accuracy and computation speed. Results show that though Pan and Tompkins shows a better accuracy (98.15%) on detecting the QRS complex compared to Derivative-based Method (96.73%), the latter consume less than half of computation time (a total 17.86 minutes to compute 24 hours ECG data). Compared to Pan and Tompkins algorithm (a total 45.2 minutes to compute 24 hours ECG data).
Keywords :
data analysis; electrocardiography; feature extraction; field programmable gate arrays; ECG QRS detection method; ECG data; ECG device; ECG system; FPGA platform; MIT-BIH database; Pan and Tompkins algorithm; adaptive threshold; benchmarking; derivative-based method; electrocardiogram QRS complex detection algorithm; feature extraction accuracy; field programmable gate array platform; fixed threshold; heart monitoring; software computation; time 17.86 min; time 24 hour; time 30 min; time 45.2 min; Accuracy; Databases; Diseases; Electrocardiography; Equations; Heart; Noise;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Engineering and Sciences (IECBES), 2014 IEEE Conference on
Type :
conf
DOI :
10.1109/IECBES.2014.7047494
Filename :
7047494