Development of a gamma radiation imaging detector based on a GSO crystal scintillator and a position sensitive PMT
Author :
Majewski, S. ; Kross, B. ; Majewski, L. ; Pohl, M. ; Steinbach, D. ; Weisenberger, A.G. ; Wojcik, R.
Volume :
2
fYear :
1996
Firstpage :
1191
Abstract :
An imaging gamma detector was developed using a 5" diameter Hamamatsu R3292 position sensitive PMT and GSO crystal scintillator plates from Hitachi. Plate sizes from 4×4 cm to 8×8 cm and thickness from 1.25 mm to 2.5 mm were used to detect 28-35 keV photons from 125I and 140 keV gamma\´s from 99mTc. Several novel features of the detector were investigated including improvement of detector resolution by using anode wire sections and truncated center-of-gravity calculations. After the PSPMT\´s readout was optimized a resolution of about ~5 mm @ 140 keV was obtained using only fourteen (7x,7y) anode wire sections in lieu of the standard 28×28 individual crossed-wires. Due to good obtained overall energy resolution, efficient scatter rejection is expected by placing just one simple energy window on scintillation signals coming from the detector. Measurements were made with point and linear sources as well as thyroid and compressed breast phantoms with simulated lesions. The phantoms contained appropriate concentrations of 99mTc emitting 140 keV gamma radiation. The detector combines good performance with small size and economical makeup. After additional phantom tests and software improvement we are planning to use it first as a small field-of-view small animal gamma imager
Keywords :
gamma-ray apparatus; medical image processing; position sensitive particle detectors; radioisotope imaging; solid scintillation detectors; GSO crystal scintillator; GSO crystal scintillator plates; Gd2(SiO4)O:Ce; Hamamatsu R3292 position sensitive PMT; anode wire sections; compressed breast phantoms; efficient scatter rejection; energy resolution; gamma radiation imaging detector; imaging gamma detector; position sensitive PMT; simulated lesions; small field-of-view small animal gamma imager; thyroid; truncated center-of-gravity calculations; Anodes; Energy resolution; Gamma ray detection; Gamma ray detectors; Gamma rays; Imaging phantoms; Optical imaging; Radiation detectors; Solid scintillation detectors; Wire;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-3534-1
Type :
conf
DOI :
10.1109/NSSMIC.1996.591620
Filename :
591620
Link To Document :
بازگشت