DocumentCode :
3567487
Title :
Using a hybrid pixel detector for dynamic radiography
Author :
Pedersen, Frands ; Sandborg, M.
Author_Institution :
Dept. of Radiat. Sci., Uppsala Univ., Sweden
Volume :
2
fYear :
1996
Firstpage :
1239
Abstract :
The hybrid pixel detector has recently been introduced in digital radiography. A Monte Carlo simulation was performed with the purpose of comparing the photon counting and energy integration registration strategies, using the detective quantum efficiency DQE. The results were compared with the input screen of an image intensifier. A cascaded model of the hybrid pixel detector is outlined
Keywords :
Monte Carlo methods; diagnostic radiography; image intensifiers; medical diagnostic computing; photon counting; position sensitive particle detectors; Monte Carlo simulation; detective quantum efficiency; digital radiography; dynamic radiography; energy integration registration; hybrid pixel detector; image intensifier; input screen; photon counting; Displays; Frequency; Image intensifiers; Radiation detectors; Radiography; Sensor arrays; Spatial resolution; X-ray detection; X-ray detectors; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-3534-1
Type :
conf
DOI :
10.1109/NSSMIC.1996.591634
Filename :
591634
Link To Document :
بازگشت