DocumentCode :
3567534
Title :
Performance of a new 3D-only PET scanner-the EXACT3D
Author :
Spinks, T.J. ; Bailey, D.L. ; Bloomfield, P.M. ; Miller, M. ; Murayama, H. ; Jones, T. ; Jones, W. ; Reed, J. ; Newport, D. ; Casey, M.E. ; Nutt, R.
Author_Institution :
MRC Cyclotron Unit, Hammersmith Hospital, London, UK
Volume :
2
fYear :
1996
Firstpage :
1275
Abstract :
Characteristics of currently the highest sensitivity 3D-only PET scanner (CTI/Siemens 966/EXACT3D) have been studied. The device has 48 rings (82 cm diameter) of BGO detector elements (4.0×4.1×30 mm, 8×8 block) and an axial FOV of 23.4 cm. Attenuation correction is carried out with a point source of 137Cs which moves under hydraulic pressure in a helical tube. Transaxial resolution is 4.7 mm 1 cm from the centre of the FOV increasing, in the radial direction to 6.3 mm and 7.3 mm at 10 cm and 15 cm. At a lower threshold of 350 keV, the scatter fraction is 40%. With scatter subtracted, the efficiency (20 cm cylinder) is 6.9×104 cps/kBq/ml (maximum ring difference of 40); the absolute sensitivity is 5.8%. In the current configuration, the maximum total coincidence event rate is limited to about 3×106 per sec. The maximum true rate is about 850 kcps (90 MBq in the FOV). List mode acquisition has been implemented to maximise temporal resolution and optimise data storage. The more open geometry of this device gives a broader singles FOV. Hence administered doses and/or additional side-shielding need to be carefully considered to optimise noise-equivalent counts
Keywords :
biomedical equipment; bismuth compounds; data acquisition; positron emission tomography; solid scintillation detectors; 3D-only PET scanner; BGO detector elements; Bi4(GeO4)3; EXACT3D; administered doses; attenuation correction; list mode acquisition; maximum total coincidence event rate; noise-equivalent counts; side-shielding; transaxial resolution; Attenuation; Cyclotrons; Detectors; Electromagnetic scattering; Geometry; Hospitals; Memory; Particle scattering; Positron emission tomography; Subspace constraints;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-3534-1
Type :
conf
DOI :
10.1109/NSSMIC.1996.591679
Filename :
591679
Link To Document :
بازگشت