Abstract
New training methods and exercises for upper limbs rehabilitation can be developed using robotics and virtual reality technologies. The technologies can make quantitative evaluations and enhance the qualitative effect of trainings. We have developed a 3D exercise machine for upper limbs (EMUL) with special actuators using electrorheological (ER) fluid. The ER actuator is composed of an ER clutch and its driven mechanism consisting of a motor and a reduction-gear unit, and we can ensure safety of patients mechanically by using them. This paper deals with the development of EMUL. We also present the development of software for motion exercise trainings and some results of clinical evaluation.
Keywords
biomechanics; electrorheology; medical robotics; patient rehabilitation; 3-D rehabilitation system; clinical testing; electrorheological fluid; reduction-gear unit; robotics; upper limbs rehabilitation; virtual reality technologies; Actuators; Aging; Biomedical engineering; Educational institutions; Erbium; Industrial training; Rehabilitation robotics; Safety; System testing; Virtual reality;
fLanguage
English
Publisher
ieee
Conference_Titel
Rehabilitation Robotics, 2005. ICORR 2005. 9th International Conference on
Print_ISBN
0-7803-9003-2
Type
conf
DOI
10.1109/ICORR.2005.1501050
Filename
1501050
Link To Document