Abstract
In this paper, an integrated loading model of ultrasonic transducer is established to explain the phenomenon that different tissue loads have different effect on resonance frequency of ultrasonic transducer. Mechanical loss of ultrasonic transducer is taken into consideration in the model and mass-spring-damping tissue model and contact impedance between ultrasonic transducer and tissues are given. Experiments show that this model is accurate enough to analyze loading characteristic of ultrasonic transducer. Effect of tissue elastic module on resonance frequency of ultrasonic transducer is studied. Results show that soft tissue with low elastic module make resonance frequency of ultrasonic transducer lower than unloaded condition, while tissue with large elastic module make resonance frequency larger.
Keywords
biomedical transducers; ultrasonic transducers; integrated loading model; loading mechanism; mass-spring-damping tissue model; ultrasonic biomedical transducer; Admittance; Admittance measurement; Bones; Integrated circuit modeling; Load modeling; Resonant frequency; Ultrasonic transducers; contact impedance; loading characteristic; loss; tissue model; ultrasonic transducer;
fLanguage
English
Publisher
ieee
Conference_Titel
Computer Science and Electronics Engineering (ICCSEE), 2012 International Conference on
Print_ISBN
978-1-4673-0689-8
Type
conf
DOI
10.1109/ICCSEE.2012.258
Filename
6188078
Link To Document