DocumentCode
3581580
Title
Doppler radar for sleep medicine
Author
Baboli, Mehran ; Soll, Bruce ; Boric-Lubecke, Olga ; Lubecke, Victor
Author_Institution
Department of Electrical Engineering, University of Hawaii at Manoa, USA
fYear
2014
Firstpage
953
Lastpage
955
Abstract
A design for a physiological radar monitoring system (PRMS) is presented that can be integrated with clinical sleep monitoring systems. The PRMS uses two radar systems at 2.45 GHz and 24 GHz to achieve both high sensitivity and high resolution. The system can acquire data, perform digital processing and output appropriate conventional analog outputs, which can be read and interpreted in a standard sleep monitoring system. The PRMS system was tested on ten patients and the results compared favorably with a gold standard monitoring system. The results show the possibility of integrating a PRMS with a standard monitoring system to achieve more informative results.
Keywords
Biomedical monitoring; Doppler radar; Monitoring; Sleep apnea; Standards; Doppler radar; Obstructive Sleep Apnea (OSA); Sandman; physiological radar monitoring system (PRMS);
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Conference (APMC), 2014 Asia-Pacific
Type
conf
Filename
7067589
Link To Document