Title :
Development of measurement equipment for calibrating beam position monitor in HLS storage ring
Author :
Shen, Lianguan ; Wang, Junhua ; Wang, Guicheng ; He, Xin ; Zhao, Jianbin ; Zhu, Yangbin ; Yao, Jianping
Author_Institution :
Dept. of Precision Machinery & Precision Instrum., Univ. of Sci. & Technol. of China, Hefei, China
fDate :
6/23/1905 12:00:00 AM
Abstract :
An automatic calibration system for the beam position monitor (BPM) of the HLS (Hefei Synchrotron Radiation Light Source) storage ring is described. The system integrated a manipulator for holding a simulated antenna vertically to move in horizontal plane, a signal collector and a data processor. The positioning accuracy of an antenna, thin wire, is monitored by a set of microscopes with reading resolution of 0.1 μm. Optical gratings with a veracity of 5 μm give the position of the antenna. The system has been used for the 2nd phase of HLS to calibrate the BPM detectors and measured 4 vacuum chambers, which have been installed in the storage ring. Test results show that the performance of the system is good enough for calibration
Keywords :
calibration; manipulators; particle beam diagnostics; position measurement; HLS storage ring; beam position monitor; calibration system; manipulator; optical gratings; positioning accuracy; vacuum chamber; Antenna measurements; Calibration; Computerized monitoring; High level synthesis; Light sources; Position measurement; Radiation monitoring; Signal processing; Storage rings; Synchrotron radiation;