Abstract :
A. first generation 120 micron pitch pixel array system for neutron detection using the PATARA amplifier chip was assembled and tested. The pixel array was tested for neutron response and spatial resolution. Pulses from the PATARA were observed at 0.5 V in height and 500 ns wide from neutron interactions. The spatial resolution of the array was determined to be 119 micrometers. Leakage current tests and alpha particle irradiation tests were conducted for a second generation prototype silicon sensor with 175 micrometer deep perforated trench structures in each pixel. The second generation sensor incorporates several design improvements to ease fabrication.
Keywords :
alpha-particle detection; leakage currents; neutron detection; nuclear electronics; readout electronics; silicon radiation detectors; PATARA amplifier chip; alpha particle irradiation; leakage current; neutron detection; neutron scattering; prototype silicon sensor; readout electronics; silicon pixel array; Alpha particles; Assembly systems; Fabrication; Leakage current; Neutron spin echo; Prototypes; Pulse amplifiers; Silicon; Spatial resolution; System testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
ISSN :
1095-7863
Print_ISBN :
978-1-4244-0922-8
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2007.4436614
Filename :
4436614
Link To Document :
بازگشت