DocumentCode
3595053
Title
Precision mechanically scanned HF ultrasound arrays
Author
Tutwiler, R.L. ; Stitt, J.P. ; Ritter, T.A. ; Shung, K.K. ; Hackenberger, Wesley S. ; Jiang, X.N. ; Rehrig, P.W.
Author_Institution
Appl. Res. Lab., Pennsylvania State Univ., University Park, PA, USA
Volume
1
fYear
2003
Firstpage
873
Abstract
An actuated synthetic-aperture array (ASA) system is described that is capable of overcoming the technical difficulties in manufacturing an ultrasonic phased array in the 30 to 70 MHz frequency range. The system is capable of generating images using a linear array of transducers which are of equal quality compared to a 64-element phased array while. The array was rapidly and precisely positioned with a piezoelectric actuator. The image was formed utilizing a synthetic aperture (SA) approach rather than electronic beamforming. The SA image was further refined by application of a deconvolution algorithm.
Keywords
biomedical ultrasonics; deconvolution; piezoelectric actuators; ultrasonic transducer arrays; 30 to 70 MHz; deconvolution algorithm; electronic beamforming; high frequency ultrasound arrays; images; piezoelectric actuator; synthetic aperture array; transducer linear array; ultrasonic phased array; Frequency; Hafnium; Image generation; Manufacturing; Phased arrays; Piezoelectric actuators; Piezoelectric transducers; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics, 2003 IEEE Symposium on
Print_ISBN
0-7803-7922-5
Type
conf
DOI
10.1109/ULTSYM.2003.1293538
Filename
1293538