DocumentCode
3600069
Title
EMF changes of Pt/Pd thermocouples at the Cu freezing point
Author
Ogura, Hideki ; Numajiri, Haruhiko ; Izuchi, Masaya ; Arai, Masaru
Author_Institution
Nat. Metrol. Inst. of Japan, Nat. Inst. of Adv. Ind. Sci. & Technol., Tsukuba, Japan
Volume
1
fYear
2002
Firstpage
495
Abstract
The drift in EMF of Pt/Pd thermocouples during successive calibrations at the Cu freezing point was measured. The different heat treatment was performed to each thermocouple before the calibration. The magnitude of the drift of the thermocouples depended on the heat treatment. The heat treatment at 850°C or 1030°C for 100 hours resulted in reduction of the drift.
Keywords
calibration; copper; electric potential; freezing; heat treatment; palladium; platinum; thermocouples; 100 hr; 1030 degC; 850 degC; Cu; Pd; Pt; calibrations; copper; freezing point; heat treatment; palladium; platinum; temperature drift reduction; thermocouple; Annealing; Calibration; Furnaces; Heat treatment; Heating; Metrology; Temperature measurement; Thermal stability; Wires;