DocumentCode :
3604777
Title :
Enzyme Sensor With Polydimethylsiloxane Membrane and CMOS Potentiostat for Wide-Range Glucose Measurements
Author :
Mross, Stefan ; Furst, Peter ; Pierrat, Sebastien ; Zimmermann, Tom ; Vogt, Holger ; Kraft, Michael
Author_Institution :
Fraunhofer Inst. for Microelectron. Circuits & Syst., Duisburg, Germany
Volume :
15
Issue :
12
fYear :
2015
Firstpage :
7096
Lastpage :
7104
Abstract :
We report on an electrochemical measurement setup comprising a glucose sensor and a CMOS potentiostat with a two-layer membrane as the first steps toward the development of an integrated in-situ sensor system for bioreactors. The potentiostat has a chip size of 2.1 mm × 2.5 mm and a linear current range from -220 nA to 240 nA with a linearity of R2 = 0.9995. For wide range measurements of glucose concentrations in cell culture media, electrodes functionalized with the enzyme glucose oxidase were spin-coated with membranes made from polydimethylsiloxane (PDMS). A two-stage curing scheme of the PDMS was applied, and different membrane thicknesses and curing times were evaluated. With these membranes, glucose concentrations up to 500 mM were measured with a linear measurement range up to 200 mM. The sensors were successfully employed in the glucose monitoring of a culture of Saccharomyces cerevisiae to monitor the glucose consumption of the cells. For interference elimination, the cellulose acetate membranes were employed.
Keywords :
biomembranes; bioreactors; biosensors; electrochemical sensors; enzymes; polymers; potentiometers; sugar; CMOS potentiostat; PDMS; Saccharomyces cerevisiae; bioreactors; cell culture media; cellulose acetate membranes; curing times; electrochemical measurement setup; electrodes; enzyme glucose oxidase; enzyme sensor; glucose concentrations; glucose monitoring; glucose sensor; integrated in-situ sensor system; interference elimination; membrane thicknesses; polydimethylsiloxane membrane; two-layer membrane; two-stage curing scheme; wide-range glucose measurements; Biochemistry; Biomembranes; Curing; Current measurement; Electrodes; Sensors; Sugar; CMOS potentiostat; Glucose sensor; PDMS; bioreactor monitoring; cellulose acetate; glucose sensor; high glucose concentrations; membrane; wide range measurement;
fLanguage :
English
Journal_Title :
Sensors Journal, IEEE
Publisher :
ieee
ISSN :
1530-437X
Type :
jour
DOI :
10.1109/JSEN.2015.2470111
Filename :
7214214
Link To Document :
بازگشت