Experimental 2.4 GHz CMOS RF front-ends for non-contact vital-sign sensing
Author :
Wang, S. ; Chang, R.-H.
Author_Institution :
Dept. of Electron. Eng., Nat. Taipei Univ. of Technol., Taipei, Taiwan
Volume :
51
Issue :
23
fYear :
2015
Firstpage :
1846
Lastpage :
1848
Abstract :
A 2.4 GHz CMOS front-ends for non-contact vital-sign sensing including respiration and heartbeat is presented. The sensor using one circulator and one differential oscillator not only achieves the injection-locked function but also reduces the system complexity. Moreover, the injection-locked signals are then down-converted by a phase demodulator constituted by a lumped power divider and a mixer. Finally, the signals are bandpass filtered and amplified for subsequently performing the vital signals. Under total power consumption of 32 mW, the respiration and heartbeat information can be recognised correctly up to 1.2 m away from the subject.
Keywords :
CMOS integrated circuits; UHF integrated circuits; bioelectric phenomena; biomedical electronics; demodulators; electric sensing devices; low-power electronics; oscillators; power dividers; CMOS RF front-ends; circulator; differential oscillator; frequency 2.4 GHz; injection-locked function; lumped power divider; noncontact vital-sign sensing; phase demodulator; power 32 mW; system complexity;