DocumentCode :
3611881
Title :
Energy Dependent Efficiency in Low Background Alpha Measurements and Impacts on Accurate Alpha Characterization
Author :
Kawasaki, Hiroyoshi ; Clark, Brett M. ; Nishino, Tomoaki ; Gordon, Michael S.
Author_Institution :
Senju Metal Ind. Co. Ltd., Tochigi, Japan
Volume :
62
Issue :
6
fYear :
2015
Firstpage :
3034
Lastpage :
3039
Abstract :
A difference in low emissivity alpha particle measurements for the same samples was observed between an Alpha Sciences model 1950 gas proportional counter and the XIA UltraLo-1800 ionization chamber. The energy dependent efficiency was evaluated for each instrument to reconcile the difference, and the implication of instrument design bias on measurement results is presented.
Keywords :
alpha-particle detection; alpha-particles; accurate alpha characterization; energy dependent efficiency; low background alpha measurements; low emissivity alpha particle measurements; Alpha particles; Ionization chambers; Single event upsets; Alpha particle; emission; errors; single event upsets;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2015.2499161
Filename :
7348787
Link To Document :
بازگشت