DocumentCode :
3612555
Title :
Single Cell Mass Measurement Using Drag Force Inside Lab-on-Chip Microfluidics System
Author :
Rahman, Md Habibur ; Ahmad, Mohd Ridzuan ; Takeuchi, Masaru ; Nakajima, Masahiro ; Hasegawa, Yasuhisa ; Fukuda, Toshio
Author_Institution :
Dept. of Comput. Sci. & Eng., Univ. of Asia Pacific, Dhaka, Bangladesh
Volume :
14
Issue :
8
fYear :
2015
Firstpage :
927
Lastpage :
934
Abstract :
Single cell mass (SCM) is an intrinsic property of single cell, it arouses a great interest among scientists as cell mass depends on the synthesis of proteins, DNA replication, cell wall stiffness, cell cytoplasm density, cell growth, ribosome, and other analogous of organisms. To date, several great strides have been taken to the advancements of SCM measurement techniques. Nevertheless, more works are required to enable the technology to push frontier in deep analysis of SCM measurement, hence to elucidate intracellular properties. In this paper, we present a lab-on-chip microfluidics system for SCM measurement, related with the force required to drag a single cell and Newton\´s law of motion inside microfluidics channel. Drag force on the cell was generated by a pressure driven syringe micropump and the motion of the cell was measured using optical observation under an inverted microscope. This approach of measuring SCM was calibrated using known mass (77.3 pg) of a polystyrene particle of 5.2 μm diameter. Furthermore, we used Saccharomyces cerevisiae baker\´s yeast cells of different sizes ( 2mathchar"702D 7 μm diameter) for SCM measurement. Mass of 4.4 μm diameter of single yeast cell was measured as 2.12 pg which is in the range of previously reported single yeast cell mass (2-3 pg). In addition, we also studied the relation between SCM and single cell size. Results showed that single yeast cell mass increases exponentially with the increasing of single cell size.
Keywords :
bioMEMS; cellular biophysics; drag; fluidic devices; lab-on-a-chip; mass measurement; microchannel flow; microorganisms; DNA replication; Newton law-of-motion; SCM measurement techniques; Saccharomyces cerevisiae baker yeast cells; cell cytoplasm density; cell growth; cell wall stiffness; drag force; intracellular properties; inverted microscope; lab-on-chip microfluidic channel; optical observation; polystyrene particle; pressure driven syringe micropump; protein synthesis; ribosome; single cell mass measurement; size 4.4 mum; size 5.2 mum; Atmospheric measurements; Drag; Force; Force measurement; Microfluidics; Optical variables measurement; Particle measurements; Cell culture; drag force; microfluidic device; single yeast cell mass; single yeast cell size;
fLanguage :
English
Journal_Title :
NanoBioscience, IEEE Transactions on
Publisher :
ieee
ISSN :
1536-1241
Type :
jour
DOI :
10.1109/TNB.2015.2507064
Filename :
7362214
Link To Document :
بازگشت