DocumentCode :
3618326
Title :
CMUT ring arrays for forward-looking intravascular imaging
Author :
O. Oralkan;S.T. Hansen;B. Bayram;G.G. Yaralioglu;A.S. Ergun;B.T. Khuri-Yakub
Author_Institution :
E.L. Ginzton Lab., Stanford Univ., CA, USA
Volume :
1
fYear :
2004
fDate :
6/26/1905 12:00:00 AM
Firstpage :
403
Abstract :
The paper describes an annular CMUT ring array designed and fabricated for the tip of a catheter used for forward-looking intravascular imaging. A 64-element, 2-mm average diameter array was fabricated as an experimental prototype. A single element in the array is connected to a single-channel custom front-end integrated circuit for pulse-echo operation. In conventional operation, the transducer operates at around 10 MHz. In the collapsed regime, the operating frequency shifts to 25 MHz and the received echo amplitude is tripled. The SNR is measured as 23 dB in a 50-MHz measurement bandwidth for an echo signal from a plane reflector at 1.5 mm. We also performed a nonlinear dynamic transient finite element analysis for the transducer, and found these results to be in good agreement with experimental measurements, both for conventional and collapsed operation.
Keywords :
"Transducers","Integrated circuit measurements","Transient analysis","Catheters","Prototypes","Pulse circuits","Frequency","Bandwidth","Performance evaluation","Finite element methods"
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2004 IEEE
ISSN :
1051-0117
Print_ISBN :
0-7803-8412-1
Type :
conf
DOI :
10.1109/ULTSYM.2004.1417748
Filename :
1417748
Link To Document :
بازگشت