• DocumentCode
    3619937
  • Title

    Accuracy analysis of dissolved oxygen measurement system realized with Cypress PSoC configurable mixed signal array

  • Author

    M. Hrgetic;I. Krois;M. Cifrek

  • Author_Institution
    Fac. of Electr. Eng. & Comput., Zagreb Univ., Croatia
  • Volume
    3
  • fYear
    2005
  • fDate
    6/27/1905 12:00:00 AM
  • Firstpage
    1105
  • Abstract
    In this paper the accuracy analysis of the proposed dissolved oxygen measurement system based on the dark´s cell and the PSoC configurable mixed signal array chip is performed. In that process critical sources of errors are determined and associated errors are measured. The temperature, as the most affecting parameter of oxygen measurements, is analyzed in greater detail. Monitoring and regulation of temperature of the sample is crucial for precision oxygen concentration measurements. In order to achieve optimal accuracy level, system construction considerations are discussed and averaging algorithm and two point calibration methods are applied.
  • Keywords
    "Signal analysis","Oxygen","Semiconductor device measurement","Voltage","Chemical and biological sensors","Biosensors","Sensor arrays","Preamplifiers","Electric variables measurement","Temperature measurement"
  • Publisher
    ieee
  • Conference_Titel
    Industrial Electronics, 2005. ISIE 2005. Proceedings of the IEEE International Symposium on
  • Print_ISBN
    0-7803-8738-4
  • Type

    conf

  • DOI
    10.1109/ISIE.2005.1529078
  • Filename
    1529078