• DocumentCode
    3621036
  • Title

    Results from a first prototype of a Compton prostate probe

  • Author

    C. Lacasta;J. Bernabeu;D. Burdette;E. Chesi;N.H. Clinthorne;Y.K. Dewaraja;K. Honscheid;H. Kagan;G. Llosa;M. Mikuz;P. Modesto;W.L. Rogers;A. Studen;P. Weilhammer;L. Zhang;D. Zontar

  • Author_Institution
    IFIC, CSIC-UVEG, Valencia, Spain
  • Volume
    1
  • fYear
    2005
  • fDate
    6/27/1905 12:00:00 AM
  • Firstpage
    64
  • Lastpage
    67
  • Abstract
    A first prototype of a Compton prostate probe has been built using a stack of five 4times1 cm2, 1 mm thick silicon pad detectors as a scatter detector, surrounded by three scintillation detectors in which the absorption of the scattered photons takes place. The silicon pad dimensions are 1.4times1.4 mm which provide the required spatial resolution. The energy resolution in the silicon sensors is about 1.4 keV FWHM as determined from several gamma sources. The results obtained validate the simulation predictions that foresee an improvement over current SPECT techniques by a factor 16-40 in sensitivity and 4-5 in spatial resolution simultaneously for an intra-rectal probe built employing this concept and placed at 2 cm from the prostate
  • Keywords
    "Prototypes","Probes","Silicon","Solid scintillation detectors","Electromagnetic scattering","Particle scattering","Spatial resolution","Scintillation counters","Electromagnetic wave absorption","Single photon emission computed tomography"
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2005 IEEE
  • ISSN
    1095-7863
  • Print_ISBN
    0-7803-9221-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2005.1596208
  • Filename
    1596208