• DocumentCode
    3622752
  • Title

    Transmission line modelling for array transducer elements

  • Author

    S. Ayter

  • Author_Institution
    Acuson, Mountain View, CA, USA
  • fYear
    1990
  • fDate
    6/12/1905 12:00:00 AM
  • Firstpage
    791
  • Abstract
    An improved KLM (Krimholtz-Leedom-Matthaei) model for medical ultrasound array transducer elements is proposed. The proposed approach involves the use of the symmetric fundamental Lamb wave modes as the primary mode of operation in all the layers. The idea is justified by comparing the calculated resonance frequencies of isotropic slabs with those obtained from coupled mode theory. Transducer sensitivities can be predicted within +or-1 dB and transducer center frequencies can be estimated within +or-50 kHz with the proposed modifications. In particular, the effect of kerf depth variations on the transducer characteristics can successfully be explained and predicted.
  • Keywords
    "Transmission lines","Biomedical transducers","Ultrasonic transducers","Frequency estimation","Transmission line theory","Ultrasonic imaging","Ultrasonic transducer arrays","Resonance","Resonant frequency","Slabs"
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1990. Proceedings., IEEE 1990
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1990.171473
  • Filename
    171473