• DocumentCode
    3623185
  • Title

    The SISWICH, a detector telescope with intrinsic calibration

  • Author

    J. Friese;A. Gillitzer;H.J. Korner;J. Reinhold;M. Peter;M.R. Maier

  • Author_Institution
    Tech. Univ. Munchen, Germany
  • fYear
    1992
  • Firstpage
    61
  • Abstract
    To overcome the difficulty of calibration for PHOSWICH counters, a silicon-scintillator sandwich (SISWICH) has been built. This detector consists of a silicon PIN photodiode glued to the front of a CsI(Tl) scintillator crystal. The signal coming from the PIN diode contains the energy loss of the particle in the silicon, which is fast (nuclear counter effect), and the light output of the scintillator, which is slow. The component coming from the nuclear counter effect can be calibrated with a pulsar. Therefore, since the energy loss of a particle with known Z has been measured, its energy is known, and this can be used to calibrate the signal from the scintillator.
  • Keywords
    "Telescopes","Calibration","Diodes","Radiation detectors","Silicon","Energy loss","Energy resolution","Stability","Signal resolution","Particle measurements"
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
  • Print_ISBN
    0-7803-0884-0
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1992.301188
  • Filename
    301188