Title :
DDS Based NMR Magnetometer in Slovenian Calibration Laboratory
Author :
S. Begus;D. Fefer
Author_Institution :
Faculty of Electrical Engineering, University of Ljubljana, Tr?a?ka 25, 1000 Ljubljana, SLOVENIA. Email: samo.begus@fe.ini-lj.si
fDate :
6/1/2004 12:00:00 AM
Abstract :
Presented absorption type proton magnetometer uses direct digital synthesizer and DSP instead of marginal oscillator, therefore frequency measurement uncertainty is negligible. Frequency modulation of high frequency signal is used instead of magnetic field modulation. With the magnetometer described and the existent permanent magnet the accreditation range of the LMM could be extended
Keywords :
"Nuclear magnetic resonance","Magnetometers","Calibration","Laboratories","Absorption","Protons","Synthesizers","Digital signal processing","Oscillators","Frequency measurement"
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2004 Conference on
Print_ISBN :
0-7803-8493-8
DOI :
10.1109/CPEM.2004.305273