DocumentCode
3624686
Title
Adaptive recurrent system for noise cancellation and arrhythmia detection
Author
I. Provaznik;J. Holcik
Author_Institution
Inst. of Biomed. Eng., Brno Tech. Univ., Purkynova, Czech Republic
Volume
2
fYear
1994
Firstpage
1270
Abstract
A new LMS-based adaptive filter for preprocessing and analysis of ECG signals recorded under conditions of stress-tests is introduced. The adaptive filter was developed to cancel motion artifacts and myopotentials and to detect diagnostically significant waves in the ECG signal, e.g. arrhythmias. The new filter exploits a repetitivity of the ECG signal for optimal removing of the nonstationary noise. It uses a vector-variable-step-size LMS algorithm to adjust filter weights according to the nonstationary properties of the processed signal. The developed system was tested on real signals and on the CSE signal library, and the results are described.
Keywords
"Adaptive systems","Noise cancellation","Electrocardiography","Adaptive filters","Signal analysis","Adaptive signal detection","Motion detection","Least squares approximation","Signal processing","System testing"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
Print_ISBN
0-7803-2050-6
Type
conf
DOI
10.1109/IEMBS.1994.415427