DocumentCode
3628505
Title
Reliability of syringe infusion pump data channel
Author
Vytautas Markevicius;Dangirutis Navikas
Author_Institution
Department of Electronics Engineering, Kaunas University of Technology, Studentu str. 50, 51368, Lithuania
fYear
2008
Firstpage
733
Lastpage
738
Abstract
There is presented patient risk management aspects using syringe infusion pump (SIP) infusion management system. System hardware and software influence to patient risk is estimated. There are presented experiment results of IrDA channel reliability. Reliability assurance using CRC, data string inversion and data transfer repeat aspects were investigated. Various external infrared noise sources and influence to data transfer reliability were analyzed.
Keywords
"Reliability","Software","Software reliability","Hardware","Noise","Personnel","Risk management"
Publisher
ieee
Conference_Titel
Information Technology Interfaces, 2008. ITI 2008. 30th International Conference on
ISSN
1330-1012
Print_ISBN
978-953-7138-12-7
Type
conf
DOI
10.1109/ITI.2008.4588502
Filename
4588502
Link To Document