• DocumentCode
    3630540
  • Title

    Optimization of beam scattering system for IFJ PAN proton radiotherapy facility by Monte Carlo simulations

  • Author

    Joanna Dabrowska;Pawel Olko;Urszula Sowa;Jan Swakon

  • Author_Institution
    Henryk Niewodnicza?ski Institute of Nuclear Physics Polish Academy of Sciences, ul. Radzikowskiego 152, 31-342 Krak?w, Poland
  • fYear
    2008
  • Firstpage
    1073
  • Lastpage
    1076
  • Abstract
    The work is aimed at optimizing the 60 MeV proton beam spreading system at the IFJ PAN radiotherapy facility for the treatment of ocular melanoma tumors. Single scattering foil and double scattering foil systems have been investigated by Monte Carlo simulations. In the case of the single scattering foil system dose distributions were uniform to within ± 1 % over a 4-cm-diameter region, whereas in the case of the double scattering foil system the same uniformity was achieved in a region up to 3.6 cm in diameter, but its efficiency of dose rates was 5.7 times better. Since the total thickness of foils in the single and the double foil systems was equivalent, in both cases the energy loss of protons was approximately the same.
  • Keywords
    "Particle beams","Scattering","Protons","Nuclear physics","Telephony","Energy loss","Malignant tumors","Neoplasms","Energy measurement","Loss measurement"
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS ´08. IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4244-2714-7
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4774584
  • Filename
    4774584