DocumentCode :
3635779
Title :
Safety distance for medical equipments based on 2G and 3G mobile systems
Author :
Zeynep Akin Colak;Selcuk Helhel;İ. Bahadir Basyigit;Şükrü Özen
Author_Institution :
Elektrik-Elektronik Muhendisligi Bolumu, Muhendislik Fakultesi Akdeniz Universitesi, Antalya, TR
fYear :
2010
fDate :
4/1/2010 12:00:00 AM
Firstpage :
1
Lastpage :
3
Abstract :
In this study, two different mobile communication operators provide services in Turkey have been chosen that each operator has both 2G and 3G services. In this study, electromagnetic interference distance to medical equipments located in The Medical School Hospital of Akdeniz University sourced from mobiled phones have been examined. Through out different units in the hospital environment, 30 different measurements carried out, and deterioration in audio and visual signal reaction of devices was found to be associated with the distance to mobile phones. Electromagnetic interference, particularly of the ECG and ted EEG device was observed, and exposure begins with range 1.25m distance.
Keywords :
"Safety devices","Biomedical equipment","Electromyography","GSM","Downlink","Electrocardiography","Electroencephalography","Hospitals","Electromagnetic interference","Time division multiple access"
Publisher :
ieee
Conference_Titel :
Biomedical Engineering Meeting (BIYOMUT), 2010 15th National
Print_ISBN :
978-1-4244-6380-0
Type :
conf
DOI :
10.1109/BIYOMUT.2010.5479764
Filename :
5479764