• DocumentCode
    3635815
  • Title

    Current source design for MREIT technique and its experimental application

  • Author

    Tankut Topal;Evren Degirmenci;Rasim Boyacioglu;V. Emre Arpinar;B. Murat Ey?boglu

  • Author_Institution
    M?hendislik Fak?ltesi, Elektrik ve Elektronik M?hendisligi B?l?m?, Orta Dogu Teknik ?niversitesi
  • fYear
    2010
  • fDate
    4/1/2010 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Conductivity distribution can be imaged using Magnetic Resonance Electrical Impedance Tomography (MREIT) technique. This technique is composed of Magnetic Resonance Imaging (MRI) and Electrical Impedance Tomography (EIT) techniques and based on imaging conductor object using MRI system while applying electrical current to the conductor. In an MREIT system, current should be applied to the object in synchrony with the MR pulse sequence. In this study, a current source compatible with MRI system has been developed. The current source is designed so as to be controlled by the computer and used by an interface. The results obtained by the MREIT system, applying 20 mA current to the experimental phantom with the designed current source, are given and discussed in the study.
  • Keywords
    "Magnetic resonance imaging","Tomography","Impedance","Magnetic resonance","Conductors","Conductivity","Computer interfaces","Imaging phantoms","Radio frequency"
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering Meeting (BIYOMUT), 2010 15th National
  • Print_ISBN
    978-1-4244-6380-0
  • Type

    conf

  • DOI
    10.1109/BIYOMUT.2010.5479818
  • Filename
    5479818