• DocumentCode
    3635907
  • Title

    Modification of approximate worst-case measurement system for EMI filters

  • Author

    Jiří Dřínovský;Zdeněk Kejík;Václav Růžek;Jiří Zachar

  • Author_Institution
    Dept. of Radio Electronics, Brno University of Technology, Purkynova 118, 612 00, Czech Republic
  • fYear
    2010
  • fDate
    4/1/2010 12:00:00 AM
  • Firstpage
    91
  • Lastpage
    94
  • Abstract
    This paper deals with the approximate worst-case test method for testing the insertion loss of the EMI filters. The systems with 0.1 Ω and 100 Ω impedances are usually used for this testing. These measuring systems are required by the international CISPR 17 standard. The main disadvantage of this system is the 0.1 Ω impedance transformer. The dynamic range for the transformation from the 50 Ω, which is most common matched impedance for the measurement setups, to 0.1 Ω is very large. It is no easy to produce such transformers with this high impedance transformation ratio. These transformers have usually very narrow bandwidth. This paper discusses the alternative system with 1 Ω and 100 Ω impedances. The performance of these systems was tested on several filters and the obtained data are depicted, too. The performance comparison of several filters in several systems is also included. The performance of alternate worst-case system is discussed in the conclusion.
  • Keywords
    "Electromagnetic interference","Filters","Insertion loss","Measurement standards","Electromagnetic compatibility","System testing","Impedance measurement","Loss measurement","Frequency","Voltage"
  • Publisher
    ieee
  • Conference_Titel
    Microwave Techniques (COMITE), 2010 15th International Conference on
  • Print_ISBN
    978-1-4244-6341-1
  • Type

    conf

  • DOI
    10.1109/COMITE.2010.5481866
  • Filename
    5481866