• DocumentCode
    3639452
  • Title

    Detection of soft X-rays using semi-insulating GaAs detector

  • Author

    B. Zaťko;F. Dubecký;P. Boháček;V. Nečas;L. Ryć

  • Author_Institution
    Institute of Electrical Engineering, Slovak Academy of Sciences, Dú
  • fYear
    2010
  • Firstpage
    219
  • Lastpage
    222
  • Abstract
    This work deals with the performance of thin semi-insulating GaAs-based detectors of ionizing radiation DC-coupled to the low noise readout electronics. Fabricated detector have square shape Ti/Pt/Au Schottky blocking contact with dimension of 0.5 mm and full area AuGeNi eutectic alloy quasi-ohmic contact on the opposite site. The current-voltage characteristic and 241Am pulse-height spectra of the detectors were measured and evaluated. The best energy resolution of 6.7 % for 59.5 keV gamma photons was achieved at lowered temperature of 255 K. The detector and the first stage of preamplifier (JFET) was cooled using one stage Peltier cooler.
  • Keywords
    "Detectors","Gallium arsenide","Silicon","Plasma temperature","Preamplifiers","Temperature measurement","JFETs"
  • Publisher
    ieee
  • Conference_Titel
    Advanced Semiconductor Devices & Microsystems (ASDAM), 2010 8th International Conference on
  • Print_ISBN
    978-1-4244-8574-1
  • Type

    conf

  • DOI
    10.1109/ASDAM.2010.5667023
  • Filename
    5667023