  • Abstract
    The ultra low expansion cavities play a crucial role in laser stabilization, and they are essential in ion clocks. We present a method to monitor the temperature distance changes in the Fabry-Perot cavity with a spacer from an ultra low expansion material with sub-nanometer resolution. The FPC was placed into a temperature-stabilized stainless steel chamber.
  • Keywords
    "Ultrafast optics","Optical variables control","Optical resonators","Temperature measurement","Optical refraction","Resonant frequency","Cavity resonators"