• DocumentCode
    3646143
  • Title

    SAR levels in microwave breast imaging: 3-D safety assessment with plane-wave illumination

  • Author

    Adam Santorelli;Michael Vander Schueren;Milica Popovic

  • Author_Institution
    Department of Electrical and Computer Engineering, McGill University, Montreal, Canada
  • fYear
    2011
  • Firstpage
    481
  • Lastpage
    484
  • Abstract
    In this work, we present results of simulation analysis to assess the safety of the microwave-based breast cancer detection techniques. In our numerical experiments, a detailed, three-dimensional anatomical model, derived from magnetic resonance images, has been exposed to plane waves from different angles and at varying frequencies. These simulations result in profiles of specific absorption rate (SAR), a quantity that is used to measure the compliance with safety standards. Our analysis implies that the power levels needed for the functioning of microwave breast cancer detection systems remains safe in terms of acceptable SAR levels.
  • Keywords
    "Breast","Microwave imaging","Microwave theory and techniques","Numerical models","Safety","Electromagnetic heating"
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference Proceedings (APMC), 2011 Asia-Pacific
  • Print_ISBN
    978-1-4577-2034-5
  • Type

    conf

  • Filename
    6173791