• DocumentCode
    3648144
  • Title

    Real-Time Measurements of Atmospheric CO Using a Continuous-Wave Room Temperature Quantum Cascade Laser Spectrometer

  • Author

    Jingsong Li;Uwe Parchatka;Rainer Königstedt;Horst Fischer

  • Author_Institution
    Dept. of Atmos. Chem., Max Planck Inst. for Chem., Mainz, Germany
  • fYear
    2012
  • fDate
    5/1/2012 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A compact mid-infrared laser spectrometer based-on a thermoelectrically (TE) cooled continuous-wave room temperature (RT) quantum cascade laser working at a wavelength of 4.566 μm and TE-cooled detectors has been newly developed to demonstrate the applicability of high precision measurements of atmospheric CO. Wavelength modulation spectroscopy technique combined with a multiple pass absorption cell was employed to achieve high sensitivity. The Allan variance technique analysis indicates a 1-s CO measurement precision of 1.03 ppbv, improving to a minimum of 0.28 ppbv at the optimum integration time of 53 s. Long-term performance of the instrument was examined with alternate measurements of a sample gas with known concentration and ambient air at a typical 1-Hz sampling rate and a 1-hourly calibration cycle. This completely thermoelectrically cooled system shows the capability of long-term, unattended and continuous operation at RT without any complicated cryogenic cooling.
  • Keywords
    "Quantum cascade lasers","Atmospheric measurements","Standards","Laser beams","Absorption","Instruments","Calibration"
  • Publisher
    ieee
  • Conference_Titel
    Photonics and Optoelectronics (SOPO), 2012 Symposium on
  • ISSN
    2156-8464
  • Print_ISBN
    978-1-4577-0909-8
  • Electronic_ISBN
    2156-8480
  • Type

    conf

  • DOI
    10.1109/SOPO.2012.6271103
  • Filename
    6271103