• DocumentCode
    3659447
  • Title

    Design and implementation of multipurpose single channel bio signal amplifier

  • Author

    Sandesh R.S.;Nithya Venkatesan

  • Author_Institution
    SENSE, VIT University, Vandalur, Kelambakkam Road, Chennai, TamilNadu-127, India
  • fYear
    2015
  • Firstpage
    149
  • Lastpage
    153
  • Abstract
    In this study the authors have designed a multipurpose single channel bio signal amplifier for recording of ECG/EEG/EMG for clinical applications. To accomplish this we have used only three dry electrodes along with simple EEG leads to obtain the above Bio Potential signals. The authors have used the amplifiers INA 217 which has low noise and OP-07C. The results indicate that the signal extracted is very much free from noise. Furthermore the performance analysis of this pre amplifiers has been carried out and shows low Noise of 1.69nV/. Also the outcome gives the SNR approximately 57.79 dB with a very high input impedance of 711.11KΩ and low output impedance 150Ω.
  • Keywords
    "Electroencephalography","Electrodes","Electrocardiography","Signal to noise ratio","Electromyography","Immune system"
  • Publisher
    ieee
  • Conference_Titel
    Advances in Computing, Communications and Informatics (ICACCI), 2015 International Conference on
  • Print_ISBN
    978-1-4799-8790-0
  • Type

    conf

  • DOI
    10.1109/ICACCI.2015.7275600
  • Filename
    7275600