• DocumentCode
    3673711
  • Title

    Prototype of a luxmeter with high sensitivity suitable for long-term data recording

  • Author

    Roman Hrbac;Vaclav Kolar;Tomas Novak

  • Author_Institution
    Department of Electrical Engineering, VSB - Technical University of Ostrava, 17. listopadu 15, 708 33, Czech Republic
  • fYear
    2015
  • Firstpage
    71
  • Lastpage
    74
  • Abstract
    This This paper describes the development and testing of HK-01 Data Logger luxmeter. This luxmeter is able to operate automatically and record obtained data. This device is used to measure illuminance and it is characterized by a wide range, from 10mlx to 100klx. Its resolution is 1mlx. This device makes possible both manual and automatic range setting, as well as recording obtained values into the luxmeter´s internal memory, either automatically in set time intervals, or manually. Recorded data can be viewed on the display, or exported into a computer. Furthermore, this paper describes the testing of this luxmeter in a photometric laboratory, looking especially at its directional pattern and spectral qualities. The testing was carried out with the use of JETI specbox 1211 professional spectroradiometer.
  • Keywords
    "Light sources","Switches","Operational amplifiers","Sensitivity","Microcontrollers","Battery charge measurement","Testing"
  • Publisher
    ieee
  • Conference_Titel
    Applied Electronics (AE), 2015 International Conference on
  • ISSN
    1803-7232
  • Print_ISBN
    978-8-0261-0385-1
  • Type

    conf

  • Filename
    7301059