• DocumentCode
    3673713
  • Title

    Wrist cuff method determines diastolic pressure in dual cuff blood pressure system

  • Author

    Jiri Jilek;Milan Stork

  • Author_Institution
    Carditech, Culver City, California, USA
  • fYear
    2015
  • Firstpage
    85
  • Lastpage
    88
  • Abstract
    Single cuff noninvasive blood pressure (BP) measurements performed manually or automatically are subject to errors. We developed an experimental system that uses two cuffs in order to improve accuracy of noninvasive BP measurement. This study concentrated on a novel determination of diastolic BP using wrist cuff pulse waveforms and arm cuff pressures. The method was compared with the oscillometric method using the British Hypertension Society agreement grading system. The resulting grade was B- “good agreement”. The wrist cuff method may prove to be better than single cuff methods because it is based on physiological principles.
  • Keywords
    "Wrist","Biomedical monitoring","Blood pressure","Hypertension","Pressure measurement","Software","Manuals"
  • Publisher
    ieee
  • Conference_Titel
    Applied Electronics (AE), 2015 International Conference on
  • ISSN
    1803-7232
  • Print_ISBN
    978-8-0261-0385-1
  • Type

    conf

  • Filename
    7301062