• DocumentCode
    3675899
  • Title

    Design considerations for dipole for head MRI at 10.5T

  • Author

    Jinfeng Tian;Russell L. Lagore;J. Thomas Vaughan

  • Author_Institution
    Center for Magnetic Resonance Research, University of Minnesota, Minneapolis MN 55455, United States of America
  • fYear
    2015
  • Firstpage
    99
  • Lastpage
    100
  • Abstract
    As the world´s first 10.5T whole body human MRI scanner is ready for discoveries on new research frontiers, RF challenges need to be overcome to fully realize the benefits of the ultra-high MRI system. Dipole antennas were investigated with the Finite-Difference Time-Domain (FDTD) method. For best head MRI at 10.5T, results showed that dipole arrays should be long and conformal to the head geometry. A bottom ring connecting all dipole elements with adjustable capacitance can be helpful in improving RF homogeneity. RF and dielectric shielding, and dielectric pads around head may result in a dramatic rise of the Specific Absorption Rate (SAR) within head tissues.
  • Keywords
    "Radio frequency","Dielectrics","Magnetic resonance imaging","Magnetic heads","Dipole antennas","Arrays","Finite difference methods"
  • Publisher
    ieee
  • Conference_Titel
    RF and Wireless Technologies for Biomedical and Healthcare Applications (IMWS-BIO), 2015 IEEE MTT-S 2015 International Microwave Workshop Series on
  • Type

    conf

  • DOI
    10.1109/IMWS-BIO.2015.7303796
  • Filename
    7303796