• DocumentCode
    3684061
  • Title

    On the use of test gases of various radii to investigate molecular sieving in leak channels

  • Author

    William W. Lim;Martin P. Bucknall;Lewis Adler;David R. McKenzie;Gregg J. Suaning

  • Author_Institution
    Graduate School of Biomedical Engineering, University of New South Wales, Sydney, Australia
  • fYear
    2015
  • Firstpage
    813
  • Lastpage
    816
  • Abstract
    Evidence of the effect of molecule size (molecular sieving) was discovered in leak channels similar to those found in hermetically sealed implantable bionics. A range of test gases of different molecular sizes was used to investigate the relative leak rates of several different samples. A contemporary model of molecular sieving is shown to be in partial agreement with our data.
  • Keywords
    "Valves","Helium","Xenon","Mathematical model","Flanges","Australia"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7318486
  • Filename
    7318486