DocumentCode
3684192
Title
A smart pill for drug delivery with sensing capabilities
Author
R. Goffredo;D. Accoto;M. Santonico;G. Pennazza;E. Guglielmelli
Author_Institution
Biomedical Robotics and Biomicrosystems Lab, Università
fYear
2015
Firstpage
1361
Lastpage
1364
Abstract
In this paper a novel system for local drug delivery is described. The actuation principle of the micropump used for drug delivery relies on the electrolysis of a water-based solution, which is separated from a drug reservoir by an elastic membrane. The electrolytically produced gases pressurize the electrolytic solution reservoir, causing the deflection of the elastic membrane. Such deflection, in turn, forces the drug out of its reservoir through a nozzle. The proposed system is integrated in a swallowable capsule, equipped with an impedance sensor useful to acquire information on the physiological conditions of the tissue. Such information can be used to control pump activation.
Keywords
"Impedance","Impedance measurement","Drugs","Reservoirs","Electrodes","Drug delivery","Electrical resistance measurement"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7318621
Filename
7318621
Link To Document