    The paper presents analysis of the generic transfer function (TF) between Electrical Bioimpedance (EBI) measured non-invasively on the wrist and Central Aortic Pressure (CAP) invasively measured at the aortic root. Influence of the Heart Rate (HR) variations on the generic TF and on reconstructed CAP waveforms is investigated. The HR variation analysis is provided on a single patient data to exclude inter-patient influences at the current research stage. A new approach for the generic TF estimating from a data ensemble is presented as well. Moreover, an influence of the cardiac period beginning point selection is analyzed and empirically optimal solution for its selection is proposed.