DocumentCode
3685025
Title
Development of Stewart platform type ankle-foot device for trip prevention support
Author
Kenta Nomura;Teru Yonezawa;Takeki Ogitsu;Hiroshi Mizoguchi;Hiroshi Takemura
Author_Institution
Tokyo University of Science, 2641, Yamazaki, Noda-shi, Chiba, Japan
fYear
2015
Firstpage
4808
Lastpage
4811
Abstract
This paper presents an ankle-foot device using a Stewart platform, which is a type of parallel-link mechanism, for trip prevention support. The developed device can reproduce the input motions of the ankle joint in six degrees of freedom by controlling six pneumatic cylinders at the same time. The root mean square errors of the 3-D position and rotation angle of the reproduced motions with the input motions (dorsiflexion and plantar flexion) were 6.3 mm and 3.0°, respectively. Verification experiments for trip prevention support performance were conducted by comparing motions in each walking condition measured using a motion capture system. The experimental results showed that the minimum foot clearance during mid-swing and initial swing increased significantly by the trip prevention support offered by the developed device. The developed device can perform passive exercises for ankle rehabilitation and support walking for trip prevention.
Keywords
"Legged locomotion","Foot","Performance evaluation","Joints","Senior citizens","Sociology","Statistics"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7319469
Filename
7319469