  • Abstract
    This paper proposes a new master device for teleoperated needle insertion-type interventional robotic system. The 5-DOF master device is optimally designed based on the newly defined interventional procedures and the physicians´ requirements. It comprises a 2-DOF rotational mechanism for adjustment of needle orientation, a 2-DOF translational mechanism for fine-tuning of needle entry point, and a handle assembly. The handle assembly includes a 1-DOF translational mechanism for needle insertion and buttons for operation mode selection. The passive actuation modules of the rotational mechanism and the active actuation modules of the translational mechanism are controlled appropriately for the selected mode according to the procedure phase. The needle insertion mechanism also warns the user by vibrating the shaft when the needle reaches the dangerous region.