DocumentCode
3685036
Title
Design optimization of neuroendoscopic continuum instruments for third ventriculostomy and tumor biopsy
Author
Kyle W. Eastwood;Thomas Looi;Hani E. Naguib;James M. Drake
Author_Institution
Institute for Biomaterials and Biomedical Engineering, University of Toronto, ON M5S 2J7 CAN
fYear
2015
Firstpage
4853
Lastpage
4856
Abstract
A simulation-based approach to the design of procedure-specific dexterous neuroendoscopic continuum instruments for endoscopic third ventriculostomy and tumor biopsy is proposed. Given pre-operative CT and MRI images, the algorithm returns instrument design specifications including lengths, curvatures and alternative positions for the surgical incision while respecting anatomical boundaries. This methodology proposes a novel clinically-guided geometric representation of surgical targets specific to this indication. The additional constraints imposed by the presence of surgical trocars and endoscopes are also considered. Finally, we conclude with a clinical example to demonstrate the proposed approach.
Keywords
"Instruments","Robots","Tumors","Neurosurgery","Optimization","Biopsy"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7319480
Filename
7319480
Link To Document