  • Title

    Design optimization of neuroendoscopic continuum instruments for third ventriculostomy and tumor biopsy

  • Author

    Kyle W. Eastwood;Thomas Looi;Hani E. Naguib;James M. Drake

  • Author_Institution
    Institute for Biomaterials and Biomedical Engineering, University of Toronto, ON M5S 2J7 CAN
  • fYear
    2015
  • Firstpage
    4853
  • Lastpage
    4856
  • Abstract
    A simulation-based approach to the design of procedure-specific dexterous neuroendoscopic continuum instruments for endoscopic third ventriculostomy and tumor biopsy is proposed. Given pre-operative CT and MRI images, the algorithm returns instrument design specifications including lengths, curvatures and alternative positions for the surgical incision while respecting anatomical boundaries. This methodology proposes a novel clinically-guided geometric representation of surgical targets specific to this indication. The additional constraints imposed by the presence of surgical trocars and endoscopes are also considered. Finally, we conclude with a clinical example to demonstrate the proposed approach.