DocumentCode :
3685101
Title :
Intraoperative monitoring of intestinal viability: Evaluation of a new combined sensor
Author :
Zahra McGuinness-Abdollahi;Mohamed A. Thaha;Anthony Ramsanahie;Shafi Ahmed;Panayiotis A. Kyriacou;Justin P. Phillips
Author_Institution :
School of Mathematics, Computer Science and Engineering, City University London, EC1V 0HB, UK
fYear :
2015
Firstpage :
5126
Lastpage :
5129
Abstract :
A dual wavelength photoplethysmography (PPG) and laser Doppler flowmetry (LDF) sensor was developed to investigate the suitability of these techniques for monitoring bowel viability intraoperatively. Clinical measurements were obtained from thirty patients undergoing bowel surgery. Three measurements were performed at different stages of the operation. The amplitude of infrared PPG decreased from the baseline measurement to the pre-anastomosis measurement by 36% and LDF flux decreased by 21% for the same measurements. An increase of 33% in amplitude for infrared PPG was observed from the pre-anastomotic to post-anastomosis measurement; the equivalent increase was not seen for LDF flux. The results revealed that the sensor could potentially indicate changes in perfusion and blood flow at critical phases of surgery, thereby assisting in the early detection of inadequate blood supply in bowel tissue. The results also suggest that laser Doppler is more sensitive to movement artefact compared to PPG.
Keywords :
"Surgery","Probes","Doppler effect","Monitoring","Cancer","Blood flow"
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN :
1094-687X
Electronic_ISBN :
1558-4615
Type :
conf
DOI :
10.1109/EMBC.2015.7319545
Filename :
7319545
Link To Document :
بازگشت