• DocumentCode
    3685236
  • Title

    Carotid arterial blood pressure waveform monitoring using a portable ultrasound system

  • Author

    Joohyun Seo;Sabino J. Pietrangelo;Hae-Seung Lee;Charles G. Sodini

  • Author_Institution
    Microsystems Technology Laboratories, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, 02139, USA
  • fYear
    2015
  • Firstpage
    5692
  • Lastpage
    5695
  • Abstract
    This work presents a non-invasive arterial blood pressure (ABP) waveform monitoring technique using ultrasound. A portable ultrasound system to excite ultrasound transducers and acquire data is designed with off-the-shelf components. The insonation angles are identified using a vector Doppler technique based on the cosine dependency of the Doppler signals. The pulse pressure of an estimated waveform at the left common carotid artery is compared to the standard sphygmomanometer measurement in a clinical test. The estimated carotid ABP waveform shows excellent agreement to the finger ABP waveform with expected discrepancy of the systolic peak shape due to different measurement sites. The proposed method also tracks slow blood pressure fluctuations. This validation on human subjects shows potential for a noninvasive blood pressure waveform monitoring device at central arterial sites.
  • Keywords
    "Ultrasonic imaging","Transducers","Biomedical monitoring","Blood pressure","Estimation","Monitoring","Doppler effect"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7319684
  • Filename
    7319684