DocumentCode :
3685412
Title :
Discrete wavelength selection for the optical readout of a metamaterial biosensing system for glucose concentration estimation via a support vector regression model
Author :
T. Teutsch;M. Mesch;H. Giessen;C. Tarín
Author_Institution :
Institute for System Dynamics, University of Stuttgart, 70569, Germany
fYear :
2015
Firstpage :
6421
Lastpage :
6424
Abstract :
In this contribution, a method to select discrete wavelengths that allow an accurate estimation of the glucose concentration in a biosensing system based on metamaterials is presented. The sensing concept is adapted to the particular application of ophthalmic glucose sensing by covering the metamaterial with a glucose-sensitive hydrogel and the sensor readout is performed optically. Due to the fact that in a mobile context a spectrometer is not suitable, few discrete wavelengths must be selected to estimate the glucose concentration. The developed selection methods are based on nonlinear support vector regression (SVR) models. Two selection methods are compared and it is shown that wavelengths selected by a sequential forward feature selection algorithm achieves an estimation improvement. The presented method can be easily applied to different metamaterial layouts and hydrogel configurations.
Keywords :
"Sugar","Metamaterials","Support vector machines","Estimation","Mathematical model","Sensors","Wavelength measurement"
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN :
1094-687X
Electronic_ISBN :
1558-4615
Type :
conf
DOI :
10.1109/EMBC.2015.7319862
Filename :
7319862
Link To Document :
بازگشت