• DocumentCode
    3685586
  • Title

    Comparison of foam-based and spring-loaded dry EEG electrodes with wet electrodes in resting and moving conditions

  • Author

    Arnold Yeung;Harinath Garudadri;Carolyn Van Toen;Patrick Mercier;Ozgur Balkan;Scott Makeig;Naznin Virji-Babul

  • Author_Institution
    University of British Columbia, Vancouver, V6T 1Z4 Canada
  • fYear
    2015
  • Firstpage
    7131
  • Lastpage
    7134
  • Abstract
    The introduction of dry electrodes for EEG measurements has opened up possibilities of recording EEG outside of standard clinical environments by reducing required preparation and maintenance. However, the signal quality of dry electrodes in comparison with wet electrodes has not yet been evaluated under activities of daily life (ADL) or high motion tasks. In this study, we compared the performances of foam-based and spring-loaded dry electrodes with wet electrodes under three different task conditions: resting state, walking, and cycling. Our analysis showed that signals obtained by the 2 types of dry electrodes and obtained by wet electrodes displayed high correlation for all conditions, while being prone to similar environmental and electrode-based artifacts. Overall, our results suggest that dry electrodes have a similar signal quality in comparison to wet electrodes during motion and may be more practical for use in mobile and real-time motion applications due to their convenience. In addition, we conclude that as with wet electrodes, post-processing can mitigate motion artifacts in ambulatory EEG acquisition.
  • Keywords
    "Electrodes","Electroencephalography","Correlation","Legged locomotion","Scalp","Mobile communication"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7320036
  • Filename
    7320036