DocumentCode :
3685627
Title :
Quantification of velocity reduction after flow diverter placement in intracranial aneurysm: An ex vivo study with 3D printed replicas
Author :
Jeff R. Anderson;Richard Klucznik;Orlando Diaz;Y. Jonathan Zhang;Gavin W. Britz;Robert G. Grossman;Christof Karmonik
Author_Institution :
MRI Core Facilities at the Houston Methodist Research Institute, TX 77030 USA
fYear :
2015
Firstpage :
7300
Lastpage :
7303
Abstract :
Phase contrast MRI (pcMRI) was used to measure flow before and after placement of a flow diverter (n = 3). Decreases from 18% to 31% in flow velocity were seen in the inflow jet of the aneurysms. Flow patterns were also compared. It was observed that the gross aneurysmal flow patterns were maintained after flow diverter placement despite decreased fluid velocities. All measurements were carried out in 3D printed aneurysm replicas.
Keywords :
"Aneurysm","Magnetic resonance imaging","Computational fluid dynamics","Three-dimensional displays","Yttrium","Neck","Neurosurgery"
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN :
1094-687X
Electronic_ISBN :
1558-4615
Type :
conf
DOI :
10.1109/EMBC.2015.7320077
Filename :
7320077