Title :
Effects of stray capacitance to ground in three electrode monopolar needle bioimpedance measurements
Author :
H. Kalvøy;C. Aliau-Bonet;R. Pallas-Areny;Ø.G. Martinsen
Author_Institution :
Clinical and Biomedical engineering department at Rikshospitalet, Oslo University Hospital, Norway
Abstract :
Positive phase angle is documented and analyzed in a three electrode monopolar needle measurement. Inductance equivalent behavior of the stray capacitance to ground is described as error source in a non-inductive sample measurement.
Keywords :
"Electrodes","Impedance","Impedance measurement","Capacitance","Needles","Frequency measurement","Biomedical measurement"
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
Electronic_ISBN :
1558-4615
DOI :
10.1109/EMBC.2015.7320137