  • Title

    Estimation of physiological sub-millimeter displacement with CW Doppler radar

  • Author

    Jia Xu;Xiaomeng Gao;Bryson E Padasdao;Olga Boric-Lubecke

  • Author_Institution
    Electrical Engineering Department, University of Hawaii at Manoa, Honolulu, 96848 USA
  • fYear
    2015
  • Firstpage
    7602
  • Lastpage
    7605
  • Abstract
    Doppler radar physiological sensing has been studied for non-contact detection of vital signs including respiratory and heartbeat rates. This paper presents the first micrometer resolution Wi-Fi band Doppler radar for sub-millimeter physiological displacement measurement. A continuous-wave Doppler radar working at 2.4GHz is used for the measurement. It is intended for estimating small displacements on the body surface resulting from physiological activity. A mechanical mover was used as target, and programmed to conduct sinusoidal motions to simulate pulse motions. Measured displacements were compared with a reference system, which indicates a superior performance in accuracy for having absolute errors less than 10μm, and relative errors below 4%. It indicates the feasibility of highly accurate non-contact monitoring of physiological movements using Doppler radar.
  • Keywords
    "Doppler radar","Displacement measurement","Cameras","Estimation","Radar measurements","Accuracy"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7320152
  • Filename
    7320152