DocumentCode
3685786
Title
Investigation of photoplethysmography and arterial blood oxygen saturation from the ear-canal and the finger under conditions of artificially induced hypothermia
Author
K Budidha;P A Kyriacou
Author_Institution
Biomedical Engineering Research Centre, School of Engineering and Mathematical Sciences, City University London, UK, EC1V 0HB
fYear
2015
Firstpage
7954
Lastpage
7957
Abstract
Pulse oximeters relay on the technique of photoplethysmography (PPG) to estimate arterial oxygen saturation (SpO2). In conditions of poor peripheral perfusion such as hypotension, hypothermia, and vasoconstriction, pulse oximeters become inaccurate or provide no reading. This is due to the poor quality of the PPG signals detected at that instance. In order to overcome this problem, the ear canal has been proposed as a alternative measurement site for measuring reliable SpO2. Hence, an ear canal PPG sensor was developed along with a PPG processing system. The performance of the sensor was evaluated by measuring the red and infrared PPGs and SpO2 from 10 healthy volunteers undergoing artificially induced hypothermia. The results from the ear canal sensor were compared with simultaneously acquired results from the finger. Hypothermia was induced by exposing the volunteers to cold temperatures of 10 ± 1°C. The results acquired suggest that the ear canal pulse oximeter endures more in estimating SpO2 values accurately when compared with the more common finger pulse oximeter.
Keywords
"Ear","Irrigation","Temperature measurement","Biomedical measurement","Probes","Electrocardiography","Blood"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7320237
Filename
7320237
Link To Document